Title
Design and construction of a prostate phantom for treatment using conformational radiotherapy
Date Issued
01 March 2012
Access level
metadata only access
Resource Type
journal article
Author(s)
Yaya G.
Pontificia Universidad Católica Del Perú (PUCP)
Abstract
Phantoms made of synthetic materials are commonly used in conformational radiotherapy to verify the proper operation of therapeutic medical equipment. The main objective of the paper is to design and build a prostate phantom to be used in the calibration of planning systems. The proposed phantom is made of nylon and acrylic. These inexpensive materials are readily available in Peru and have comparable density to soft tissue and bone. The phantom was used in the calibration of a planning system. Final results showed an average error of 5.49% between the measurements made on the phantom and planned values. These results are in the range of uncertainty recommended by the International Commission on Radiation Units and Measurements (ICRU).© 2012 IEEE.
Start page
1452
End page
1458
Volume
10
Issue
2
Language
Spanish
OCDE Knowledge area
Otros temas de medicina clínica
Subjects
Scopus EID
2-s2.0-84860267727
Source
IEEE Latin America Transactions
ISSN of the container
15480992
Sources of information:
Directorio de Producción Científica
Scopus