Title
Comparison of three ELISAs for the routine diagnosis of human T-lymphotropic virus infection in a high-prevalence setting in Peru
Date Issued
01 April 2009
Access level
metadata only access
Resource Type
journal article
Publisher(s)
Oxford University Press
Abstract
To compare three human T-lymphotropic virus (HTLV) ELISAs in a high-prevalence setting, we recruited 300 adults: 125 relatives of HTLV-infected subjects and 175 patients with possible diagnoses of HTLV-associated diseases. Sera were tested with Platelia, Murex, and Ortho ELISA. Samples with positive or discordant ELISA underwent confirmatory Inno-Lia testing. Inno-Lia gave 85/300 HTLV-1-positive and 1/300 HTLV-2-positive results. The positive predictive value was 98% for Platelia, and 100% for Murex and Ortho. Six samples had discordant ELISA; Murex gave one false-negative result, Ortho two and Platelia one. In high-prevalence settings, it is recommended to test samples with two ELISAs before considering them HTLV-seronegative. © 2008 Royal Society of Tropical Medicine and Hygiene.
Start page
420
End page
422
Volume
103
Issue
4
Language
English
OCDE Knowledge area
Tecnología para la identificación y funcionamiento del ADN, proteínas y enzimas y como influencian la enfermedad) Dermatología, Enfermedades venéreas
Scopus EID
2-s2.0-61449267420
PubMed ID
Source
Transactions of the Royal Society of Tropical Medicine and Hygiene
ISSN of the container
00359203
Sources of information: Directorio de Producción Científica Scopus