Title
LR-115 detector response to <sup>222</sup>Rn, <sup>220</sup>Rn and their progenies, exposed to hemispherical surfaces in free air, and design of a system to calculate their concentrations
Date Issued
01 August 2008
Access level
metadata only access
Resource Type
journal article
Author(s)
Universidad Simon Bolivar, Caracas,
Publisher(s)
Elsevier
Abstract
The sensitivities of LR-115 detectors to 222Rn and 220Rn chains in front of hemispherical surfaces of different radii are calculated by the Monte Carlo method. The optimum radii of hemispherical caps are determined, as a result contributions of the different α-emitter nuclei to the track density are differentiated and non-uniform track distribution is avoided. It is demonstrated that if eight detectors are exposed separately in front of the same number of hemispherical surfaces of different radii, the concentrations of each α emitter in airborne and deposited states can be determined. A passive integrating system to calculate the concentrations of radon, thoron and their progenies in free air is given. © 2008 Elsevier Ltd. All rights reserved.
Volume
43
Issue
SUPPL.1
Language
English
OCDE Knowledge area
Química inorgánica, Química nuclear
Subjects
Scopus EID
2-s2.0-48549092125
Source
Radiation Measurements
ISSN of the container
13504487
Source funding
Fondo Nacional de Ciencia Tecnología e Innovación
Sponsor(s)
The National Fund of Science, Technology and Innovation (FONACIT), Venezuela, supported this work, under Contract no. S1-2001000954.