Title
Benefits and Limitations of MARD as a Performance Parameter for Continuous Glucose Monitoring in the Interstitial Space
Date Issued
01 January 2020
Access level
open access
Resource Type
review
Author(s)
Heinemann L.
Schoemaker M.
Schmelzeisen-Redecker G.
Hinzmann R.
Kassab A.
Freckmann G.
Reiterer F.
Universidad Johannes Kepler
Publisher(s)
SAGE Publications Inc.
Abstract
High-quality performance of medical devices for glucose monitoring is important for a safe and efficient usage of this diagnostic option by patients with diabetes. The mean absolute relative difference (MARD) parameter is used most often to characterize the measurement performance of systems for continuous glucose monitoring (CGM). Calculation of this parameter is relatively easy and comparison of the MARD numbers between different CGM systems appears to be straightforward on the first glance. However, a closer look reveals that a number of complex aspects make interpretation of the MARD numbers provided by the manufacturer for their CGM systems difficult. In this review, these aspects are discussed and considerations are made for a systematic and appropriate evaluation of the MARD in clinical trials. The MARD should not be used as the sole parameter to characterize CGM systems, especially when it comes to nonadjunctive usage of such systems.
Start page
135
End page
150
Volume
14
Issue
1
Language
English
OCDE Knowledge area
Endocrinología, Metabolismo (incluyendo diabetes, hormonas)
Scopus EID
2-s2.0-85068035742
PubMed ID
Source
Journal of Diabetes Science and Technology
ISSN of the container
19322968
Sources of information: Directorio de Producción Científica Scopus