Title
Comparison of preoperative simulated and postoperative real safety distances using anterior segment OCT in patients with phakic IOL according to iris configuration
Date Issued
01 January 2022
Access level
open access
Resource Type
conference paper
Author(s)
Camino-Quezada M.
Ben-Shaul O.
Publisher(s)
International Journal of Ophthalmology
Abstract
AIM: To compare the simulated safe distance (SSD) preoperatively versus real safe distance (RSD) postoperatively in patients with iris-claw phakic intraocular lens (pIOL) implantation according to iris configuration. METHODS: Totally 60 eyes of 60 patients underwent pIOL implantation for surgical correction of myopia. Anterior chamber depth (ACD) was measured with the IOLMaster 700, and nasal and temporal safety distances (SD) were measured pre- and postoperatively using Anterior Segment Visante-OCT. SD was defined as a line measured between the edge of the optic or its simulated image to the endothelium. Eyes were divided into 3 groups: convex, concave, and plane according to preoperatory iris configuration. Statistical analysis was performed using the R program, for the comparison of independent groups and multiple comparisons, the Kruskal-Wallis test and the Dunn test were used respectively. RESULTS: Mean difference between nasal preoperative SSD and postoperative RSD was -0.36±0.38, -0.29±0.48, and -0.18±0.30 mm in the concave, convex, and plane group, respectively. Mean difference between temporal SSD and RSD was -0.36±0.37, -0.14±0.38, and -0.24±0.33 mm in the concave, convex, and plane group, respectively. There were statistically significant differences between SSD and RSS for both nasal and temporal sides in the concave and plane group (P<0.002). CONCLUSION: Preoperative SSD and postoperative RSD for iris-claw pIOL shows significant differences in patients with concave and plane iris.
Start page
741
End page
746
Volume
15
Issue
5
Language
English
OCDE Knowledge area
Oftalmología
Subjects
Scopus EID
2-s2.0-85130047666
Source
International Journal of Ophthalmology
ISSN of the container
22223959
Sources of information:
Directorio de Producción Científica
Scopus