Title
Nuclear track detector response to energetic heavy ions: study case
Date Issued
01 May 2020
Access level
metadata only access
Resource Type
journal article
Author(s)
Viloria T.
Vadillo E.
Sajo-Bohus L.
Publisher(s)
Springer Netherlands
Abstract
Nuclear track methodology (NTM), is employed to analyze PADC detectors response to different energetic ion beams. Due to the passive device linearity response to heavy ions with LET-value > 30 keV μm−1, they are employed in several fields e.g. nuclear industry, medical, environmental radiation and aircraft dosimetry. To further improve application, PADC-detectors were irradiated by beams of 4He (150 MeV nucleon−1), 12C (390 MeV nucleon−1) and 56Fe (465 MeV nucleon−1), in relatively low-dose range ≤ 1 mGy. Results provided, through etched track analysis, valuable information of interest for NTM when high LET ionizing radiation is involved.
Start page
609
End page
614
Volume
324
Issue
2
Language
English
OCDE Knowledge area
Ingeniería química
Química inorgánica, Química nuclear
Subjects
Scopus EID
2-s2.0-85083358179
Source
Journal of Radioanalytical and Nuclear Chemistry
ISSN of the container
02365731
Sources of information:
Directorio de Producción Científica
Scopus