Title
Influence of gamma irradiation on a natural source of peroxidase and its effect in the reagentless amperometric biosensor for hydrogen peroxide
Date Issued
01 January 2001
Access level
metadata only access
Resource Type
journal article
Author(s)
Instituto de Química, UNICAMP
Publisher(s)
Royal Society of Chemistry
Abstract
Peroxidase was extracted from a natural source (turnip) and irradiated with 60Co (0.1 kGy). It was then used in the construction of biosensors for hydrogen peroxide determination, in order to study the effect of the gamma irradiation on the performance of the biosensors. The biosensors were constructed using two immobilization procedures: cross-linking with glutaraldehyde and covalent binding through carbodiimide. The biosensor prepared using covalent binding through carbodiimide showed a higher sensitivity for H2O2. A good enhancement in stability and sensitivity was obtained for the biosensors from irradiated material, when compared to biosensors prepared with non-irradiated enzyme. However, the initial linear response range (1.0 to 10.0 mmol dm-3) and response time (0.5 s) were equal with or without irradiation.
Start page
739
End page
742
Volume
126
Issue
6
Language
English
OCDE Knowledge area
Bioquímica, Biología molecular
Química analítica
DOI
Scopus EID
2-s2.0-0034950316
PubMed ID
Source
Analyst
ISSN of the container
00032654
DOI of the container
10.1039/b101340h