Title
A cleaning and calibration method for the spiropro portable spirometer's pneumotachometer tube in a remote field study
Date Issued
01 April 2010
Access level
metadata only access
Resource Type
research article
Author(s)
Checkley W.
Wise R.A.
Khatry S.K.
LeClerq S.C.
West K.P.
Columbia University
Abstract
BACKGROUND: We developed a systematic method of cleaning and calibration-checking for the pneumotachometer tube of the SpiroPro portable spirometer; this method maximized spirometry accuracy in a population-based study in a remote area of Nepal.& METHODS: We tested 10 factory-calibrated pneumotachometer tubes. Each use consisted of a full set of spirometry maneuvers, per the American Thoracic Society (ATS) spirometry criteria.& RESULTS: The pneumotachometers remained accurate, per the ATS criteria, for 5-9 disinfections, but began to drift toward inaccuracy after the first disinfection. All the pneumotachometers had become inaccurate, per the ATS criteria, after 10 disinfections.& CONCLUSIONS: In a remote field setting the SpiroPro pneumotachometer tube can be cleaned and reused 5-9 times before it becomes inaccurate per the ATS criteria. Rigorous rinsing in distilled water and repeated calibration checks, at various flows up to 12 L/s, are essential for precise and accurate spirometry with the SpiroPro. Reusing the SpiroPro pneu-motachometer in a remote setting may impose measurement bias. Single use of SpiroPro pneumo-tachometers, albeit more costly, will provide better data. © 2010 Daedalus Enterprises.
Start page
443
End page
452
Volume
55
Issue
4
Language
English
OCDE Knowledge area
Química orgánica Ingeniería química
Scopus EID
2-s2.0-77954931835
PubMed ID
Source
Respiratory Care
ISSN of the container
00201324
Sources of information: Directorio de Producción Científica Scopus