Title
Insights on limits of detection, precision and accuracy in TXRF analysis of trace and major elements in environmental solid suspensions
Date Issued
01 July 2018
Access level
metadata only access
Resource Type
journal article
Author(s)
West Parana State University
Publisher(s)
Elsevier Ltd
Abstract
In this work, the analytical capability of the total reflection X-ray fluorescence technique with the S2 PICOFOX spectrometer was investigated. A set of certified reference materials was prepared as solid particulate for TXRF analysis. Experimental data of sensitivity, limits of detection and recovery for many elements were obtained. Good sensitivity and limits of detection with a good recovery range of around 90–110% were achieved. Thus, the TXRF technique exhibits a good analytical potential for its applicability on different materials.
Start page
80
End page
90
Volume
137
Language
English
OCDE Knowledge area
Farmacología, Farmacia
Ingeniería química
Subjects
Scopus EID
2-s2.0-85044442944
PubMed ID
Source
Applied Radiation and Isotopes
ISSN of the container
09698043
Sponsor(s)
We acknowledge the fellowship and financial support of the following Brazilian government agencies CAPES (grant number 23038.033201/2009-16 ), CNPq (grant number 455954/2014-3 ), and the Araucaria Foundation (grant number 18320/2011 ).